Safety Notice – Captus 4000e Thyroid Uptake Systems (serial numbers 940548–941398)

Safety Notice – Captus 4000e Thyroid Uptake Systems (serial numbers 940548–941398)

Capintec has issued a safety notice for certain Captus 4000e Thyroid Uptake Systems (serial numbers 940548–941398, manufactured between April 2019 and February 2024).

Reported issue: cases of unexpected collimator detachment (≈ 28 lbs / 12.7 kg), posing a potential risk to operators and patients.

Immediate recommendations:

  • Always lock the arm before use.
  • Handle with care, avoid abrupt movements.
  • Perform procedures with patients in a seated position rather than lying down.
  • Store the arm in a vertical position when not in use.

Corrective action:
Capintec provides a replacement kit (Hard Stop Bracket) to secure the system.
Facilities should perform a preventive inspection and contact technical support.

Support: capintecsupport@mirion.com