0

Get in Touch

Tec-Control Aluminum Breakthru Kit

The Aluminum Break­thru Kit provides a ra­pid, easy and inexpensive way to test aqueous solutions, particularly per­tech­netate generator eluate, for trace quantities of alu­minum. Aluminum forms an intense red precipitate with the indicator paper, and the intensity of the color is directly proportional to the amount of aluminum in the solution. The USP allows a concentration of aluminum ion in an injection ≤10 micrograms per milliliter (10 µg/ml) in technetium 99m eluate prepared from Molybdenum 99 formed as a result of uranium fission.

Simple Procedure:

  1. Place a drop of the eluate or solution to be tested on the indicator paper. The best procedure is to form a hanging drop using a 19-22G needle.
  2. Place a drop of the standard aluminum solution on the indicator paper. Use the same size drop.
  3. Compare the intensity of the red spot formed. If the eluate spot is less intense than the standard solution, the eluate contains less than 10 µg/ml aluminum.

Models

  • 150-780 – Chromatography Kit, Tec-Control Aluminum Breakthru Kit, 10 µg/ml
    Aluminum Breakthru Kit includes: aluminum standard, 5 ml, 10 µg/ml, 50 indicator strips and manual.
  • 150-785 – Chromatography Kit, Tec-Control Aluminum Breakthru Kit, 5 µg/ml
    Aluminum Breakthru Kit includes: aluminum standard, 5 ml, 5 µg/ml, 50 indicator strips and manual.

Frequently asked questions

What is the Tec-Control Aluminum Breakthru Kit?

Why must aluminum concentration be checked in Tc-99m generator eluate?

How does the Tec-Control Aluminum test work?

What is the maximum allowed aluminum concentration in Tc-99m eluate?

Is the Tec-Control Aluminum test fast?

Does the test require specialized equipment?

Is the test reliable and reproducible?

Should the test be performed with every Tc-99m generator elution?

Le test Tec-Control Aluminum contribue-t-il à la conformité réglementaire?