3-Part PMMA CT Phantom – head, body, pediatric

The Adult and Pediatric 3-Part PMMA CTDI Phantom is designed for CT dose measurements and quality assurance in computed tomography.

Its modular design covers several examination configurations using three nested components: pediatric head, adult head / pediatric body, and adult body. This flexibility supports CTDI evaluation across different patient profiles while maintaining a practical setup for routine QA procedures.

Manufactured from PMMA, the phantom is optimized for use with pencil-type CT ionization chambers. It supports imaging and dose requirement monitoring, CT protocol validation and constancy testing.

The set includes 13 acrylic rods for plugging unused cavities according to the selected measurement configuration.

Specifications

  • Material: PMMA
  • Phantom height: 150 mm
  • For CT dose measurements and CT quality assurance
  • Compatible with all CT scanners and pencil-type ionization chambers
  • Nested 10 cm / 16 cm / 32 cm configuration
  • 10 cm module: 5 cavities (pediatric head)
  • 16 cm module: 4 cavities (adult head / pediatric body)
  • 32 cm module: 4 cavities (adult body)
  • 13 acrylic rods for plugging unused cavities
  • Compliant with IEC 60601-2-44, IEC 61223-2-6 and IEC 61223-3-5

Models

  • VD1003105 - 3-Part PMMA CT Phantom – head, body, pediatric
  • VD0225250 - Carrying Case (with removable trolley)