CRC-55tW Dose Calibrator

The CRC-55tW Dose Calibrator from Capintec combines the functionality of a CRC-55tR Dose Calibrator with an integrated well counter, enabling both activity measurement and wipe test counting within a single system.

The dose calibrator function is based on a high-pressure deep-well ionization chamber, a proven Capintec design capable of measuring activities up to 6 Ci of Tc-99m (250 GBq) with high reliability. The chamber features a plug-and-play design for ease of use and maintenance.

The well counter function is dedicated to wipe testing and low-activity measurements. It incorporates a high-sensitivity drilled-well NaI detector, capable of measuring activities as low as 1 nCi, with counting times as short as 6 seconds. A 256-channel multichannel analyzer (MCA) provides detailed spectral analysis with manual or automatic ROI selection.

The CRC-55tW supports user-definable wipe test protocols for work areas, patient areas, unrestricted areas, and sealed source leak testing. Reporting functions allow archiving and searching of well counter data by date range, wipe type, and activity, supporting quality assurance and regulatory traceability.

All functions are controlled via an 8-inch VGA color touch screen with a menu-driven interface that is easy to learn and use in daily nuclear medicine practice.

Specifications

Display and user interface:

  • 8-inch VGA color touch screen
  • Menu-driven user interface
  • 28 programmable nuclide keys
  • Display of nuclide, activity, unit, and calibration number
  • Display of count rate values for wipe tests

Dose calibrator function:

  • High-pressure deep-well ionization chamber
  • Fill gas: ultra-pure argon at 12 atm
  • Maximum activity:
    • 6 Ci (250 GBq) Tc-99m
    • 2 Ci (74 GBq) F-18
  • Maximum resolution: 0.01 µCi (0.001 MBq)
  • Accuracy: better than ± 2 %
  • Repeatability: ± 1 % over 24 hours with continuous operation

Well counter function:

  • Detector: high-sensitivity drilled-well NaI crystal
  • Multichannel analyzer: 256-channel MCA
  • ROI selection: manual and automatic
  • Low activity measurements down to 1 nCi
  • Minimum counting time: approximately 6 s

Integrated quality control:

  • Automatic zero and background adjustment
  • Daily constancy, accuracy, and voltage tests
  • Dose calibrator QC: linearity, geometry, strip QC
  • Well counter QC: Chi-Square and MDA
  • Full system self-diagnostics

Advanced functions:

  • User-definable wipe test protocols
  • Archived well counter reports
  • Searchable results by date, type, and activity
  • Dedicated pre-set key for F-18 measurements

Functions and communications:

  • USB and PC communications
  • Compatibility with nuclear medicine management systems
  • Software upgrades via USB or flash drive
  • USB or RS-232 printer connectivity

Models

  • 5130-2216 - CRC-55tW Dose Calibrator

Optional Components:

  • 5430-0058 - Epson roll printer
  • 5430-0100 - Epson ticket printer
  • 7120-1199 - CRP-200 dose tickets & labels
  • 7300-2450 - Environmental shield
  • 7300-2004 - Chamber well insert
  • 7300-2005 - Dipper
  • 5120-2257 - Custom length liner
  • 5120-2258 - Custom length dipper

Frequently asked questions

What is the CRC-55tW Dose Calibrator?

Who is the CRC-55tW intended for?

What is the difference between the CRC-55tW and the CRC-55tR?

What types of wipe tests can be performed with the CRC-55tW?

Which quality control tests are integrated in the CRC-55tW?

Is the CRC-55tW suitable for routine use?

Is the CRC-55tW compatible with nuclear medicine IT systems?

What is the energy range of the CRC-55tW?