The 10XF-3CT is a cylindrical air ionization chamber designed for Computed Tomography Dose Index (CTDI) and Dose Length Product (DLP) measurements on CT scanners.
It provides a flat energy response across its entire active length, ensuring homogeneous and reproducible beam sampling for high-precision QA measurements.
Constructed from C552 conductive materials, the 10XF-3CT features uniform sensitivity, integrated guarding, and comes with an individual calibration certificate for traceable accuracy.
With proven stability and compatibility with IBA electrometers such as MagicMaX and DosimaX, it is the standard reference chamber for CT beam quality from 100 to 150 kV.
Specifications
• Type: air ionization chamber for CT applications
• Active volume: 3 cm³
• Active length: 100 mm
• Outer electrode diameter: 9 mm
• Materials: C552 for both electrodes
• Connector: 7-pin multi-plug connector
• Cable: 2 m flexible, low noise
• Leakage current: < ±4 × 10⁻¹⁵ A
• CT beam quality range: RQR 8 – RQR 10 (100–150 kV)
• Typical calibration factor: ND,K = 72 mGy·cm/nC (120 kV / 4.5 mm Al HVL)
• Sensitivity: 1.11 nC/Gy
• Dose measuring range: 0.015 mGy – 15 Gy
• Dose rate measuring range: 0.15 mGy/s – 0.05 Gy/s
• Dose length measuring range: 0.15 mGy·cm – 150 Gy·cm
• Uncertainty: < 5 %
• Calibration reference: RQT 9 (center of cavity volume)
• Beam incidence: central beam perpendicular to the chamber axis
Models
• VD1020115 - 10XF-3CT - Ionization Chamber