The Patient Dose Calibrator (PDC) is a reference-class instrument for field calibration of patient dose and DAP (KAP) measurement systems.
It ensures traceable and accurate results for inter-institutional dose comparisons and validation of installed DAP meters.
The PDC can be used:
- with a phantom, to simulate patient exposure and scattered radiation,
- on its elevation stand, to measure air kerma directly.
Its tough ABS housing protects the built-in ionization chambers and electronics, while the display provides DAP, DAP-rate and accumulated dose after each exposure.
Specifications
• Type: reference calibrator for dose and DAP (KAP)
• Display range DAP: 0.01 to 99 999 999 μGy·m²
• Display range air kerma: 0.001 to 99 999 999 mGy
• Digital resolution DAP: 0.01 μGy·m²
• Digital resolution dose: 0.001 mGy
• Tube voltage: 40 to 150 kV
• DAP rate (low): 1 to 1×10⁴ μGy·m²/min
• DAP rate (high): 2×10³ to 9×10⁵ μGy·m²/min
• Air kerma rate: 0.2 to 9×10³ mGy/min
• Response uniformity (50–150 kV, norm. 100 kV): ±3 %
• Equivalent filtration: 0.6 mm Al (70 kV)
• Active field size: 100×100 mm² to 300×300 mm²
• Accuracy (real conditions): ±10 %
• Accuracy (reference): ±7.5 % (10 mGy/min, 15×15 cm field, 80 kV, 2.5 mm Al)
• Automatic correction: pressure (80–106 kPa), temperature (+10 to +40 °C), humidity (10–80 %, max. 20 g/m³)
• Interface: USB for data capture / control
• Battery: Li-ion, 2 cells, > 8 h operation
• Weight: 2.32 kg
• Dimensions: 350 × 410 × 35 mm (L × W × H)
Models
• PDC - Patient Dose Calibrator – reference calibrator for patient dose and DAP (KAP)
• PDC-SS - Foam Elevation Support Stand – used for air kerma measurements
• PDC Remote Control Software (optional) – automatic data capture with customizable templates