Ionization Chamber DCT10-RS / Lemo

The DCT10-RS is a ventilated, guarded ionization chamber for CT dosimetry (CTDI/DLP). It provides a uniform response across its 100 mm active length and uses a triaxial LEMO connector for low-noise, stable readout. Supplied with a calibration certificate, it operates with the DOSIMAX plus A HV electrometer; high voltage must be enabled for ion chamber measurements.

Specifications

  • Type: ventilated, guarded ion chamber with uniform 100 mm active length
  • CT range: 100–150 kV
  • Typical calibration factor: ND,K = 72 mGy·cm/nC at 120 kV, 1.HVL 4.5 mm Al
  • Typical responsivity: 14.20 nC/Gy·cm
  • DLP range: 0.1 mGy·cm to 999 Gy·cm; DLP rate: 1.00 mGy·cm/s to 0.5 Gy·cm/s
  • Measurement uncertainty: < 5%
  • Active volume: 4.9 cm³; active length: 100 mm
  • Dimensions: outer Ø 9 mm; inner Ø 8 mm; center electrode Ø 1 mm
  • Connector: LEMO triaxial size 0; low-noise cable ~2 m (max 10 m)
  • Leakage current: < ±4×10⁻¹⁵ A
  • Beam incidence: central axis perpendicular to the chamber axis; mark towards the X-ray tube

Models

  • VD1020100 - Ionization Chamber DCT10-RS / Lemo